Comparison of Corneal Curvatures Obtained with Different Devices.
Soyoon Kim, Sung Kun Chung
Department of Ophthalmology and Visual Science, Yeouido St. Mary's Hospital, The Catholic University of Korea College of Medicine, Seoul, Korea. eyedoc@catholic.ac.kr
여러 가지 기기에서의 각막곡률 비교
김소윤⋅정성근
Department of Ophthalmology and Visual Science, Yeouido St. Mary's Hospital,
The Catholic University of Korea College of Medicine, Seoul, Korea
Abstract
PURPOSE
To evaluate device efficacy using the corneal curvature value. METHODS: Prospectively, 35 patients (70 eyes) were enrolled in the present study. Three sets of corneal curvature values were obtained using the autorefractor (RK-F1(R)), manual keratometer (OM-2(R)), partial coherence interferometry keratometer (IOL Master(R)), wavefront analyzer (KR-1W(R)), and videokeratography (Orbscan II(R)). Repeatability of each device was evaluated by coefficient of variation, standard deviation, and intraclass correlation coefficient. RM-ANOVA on ranks was used to compare the differences in corneal curvatures among the devices. The Bland-Altman plot was performed to assess measurement agreement among the devices. RESULTS: The coefficient of variation values from each device ranged from 2.92% (IOL master(R)) to 3.06% (Orbscan II(R)), and the values of intraclass correlation coefficient ranged from 0.965 (KR-1W(R)) to 0.997 (IOL master(R)). Compared with the manual keratometer, there was a maximum corneal curvature difference of 1.23 D in KR-1W(R), while the other devices had differences less than 0.82 D. CONCLUSIONS: When assessing corneal curvature, the repeatability values were similar among the 5 devices, although a difference greater than 1 D was observed when comparing the KR-1W(R) to the manual keratometer.
Key Words: Corneal curvature;Keratometry


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next