J Korean Ophthalmol Soc > Volume 56(3); 2015 > Article
Journal of the Korean Ophthalmological Society 2015;56(3):404-412.
DOI: https://doi.org/10.3341/jkos.2015.56.3.404    Published online March 15, 2015.
Comparison of Corvis ST Tonometer to Other Tonometers and Clinical Usefulness of Corvis ST Tonometer.
Jung Woo Han, Seung Joo Ha
Department of Ophthalmology, Soonchunhyang University Hospital, Soonchunhyang University College of Medicine, Seoul, Korea. sjha@schmc.ac.kr
Corvis ST 안압계와 타안압계와의 안압 비교 및 녹내장 환자에서의 임상적 유용성
한정우⋅하승주
순천향대학교 의과대학 순천향대학교병원 안과학교실
Abstract
PURPOSE
To compare the accuracy and agreement of intraocular pressure (IOP) and central corneal thickness (CCT) measurements with noncontact tonometer Corvis Scheimpflug Technology (Corvis ST) versus noncontact tonometer (NCT), Goldmann applanation tonometer (GAT), rebound tonometer (RBT), and ultrasound-based corneal pachymetry (US-CCT). The secondary objective was to evaluate the corneal biomechanical values using Corvis ST tonometer in patients with glaucoma. METHODS: Thirty-one healthy participants and 47 patients with primary open angle glaucoma and normal tension glaucoma were enrolled in this study. One eye was selected randomly. In each participant, GAT, NCT, RBT, US-CCT and measurements with Corvis ST (Corvis-IOP and Corvis-CCT) were obtained. IOP and CCT measurements of each device were compared. Device agreement was calculated by Bland-Altman analysis. Additionally, corneal highest concavity parameters were compared between healthy subjects and glaucoma patients. RESULTS: Mean IOPs in all examined eyes were 13.28 +/- 2.32 mm Hg for CST, 14.71 +/- 2.95 mm Hg for GAT, 14.44 +/- 3.10 mm Hg for NCT, and 13.23 +/- 2.89 mm Hg for RBT. There was no statistical difference in IOP measurements among tonometers. Correlation analysis showed a high correlation between each pair of tonometers (all p < 0.0001). Bland-Altman plots of all devices revealed good agreement of the IOP and CCT measurements. In glaucoma patients, highest concavity time and peak distance of highest concavity parameters were statistically lower than in normal subjects (16.93 +/- 0.66 ms vs. 16.48 +/- 0.84 ms p = 0.020, 4.23 +/- 1.34 mm vs. 3.41 +/- 1.27 mm p = 0.017, respectively). CONCLUSIONS: The CST, a newly-developed tonometer with features of visualization and measurement of the corneal deformation response to an air impulse, can be considered a reliable alternative method for measuring IOP and CCT in healthy subjects and glaucoma patients. Highest concavity parameters could be another important indicator identifying corneal viscosity or elasticity in patients with glaucoma.
Key Words: Corneal biomechanics values;Corvis scheimpflug technology tonometer;Glaucoma;Intraocular pressure


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next