Comparison of Intraocular Pressure Correction Programs in Pentacam after Corneal Refractive Surgery.
Lian Hua Hong, Min Kyu Lee, Chang Won Park, Dong Jin Chang, Ying Jun Li, Choun Ki Joo
1Catholic Institute of Visual Science, The Catholic University of Korea College of Medicine, Seoul, Korea. ckjoo@catholic.ac.kr
2Department of Ophthalmology and Visual Science, The Catholic University of Korea School of Medicine, Seoul, Korea.
3Department of Ophthalmology, Yanbian University, Jilin, China.
각막굴절교정수술 후 펜타캠을 이용한 안압 보정방법의 비교
홍련화1⋅이민규2⋅박창원1⋅장동진2⋅리영준3⋅주천기1,2
Catholic Institute of Visual Science, The Catholic University of Korea College of Medicine1, Seoul, Korea
Department of Ophthalmology and Visual Science, The Catholic University of Korea School of Medicine2, Seoul, Korea
Department of Ophthalmology, Yanbian University3, Jilin, China
Abstract
PURPOSE
To evaluate the accuracy of Pentacam(R) built-in 5 intraocular pressure (IOP) correction programs used to measure the IOP of patients who received corneal refractive surgery. METHODS: IOP of 124 eyes from 62 patients who underwent epipolis laser in situ keratomileusis was measured with Goldmann applanation tonometry (GAT) at 6 months pre- and post-operatively. The collected data was input into Pentacam(R), calculated by 5 correction programs, Ehlers, Shah, Dresden, Orssengo / Pye, Kohlhaas, and compared. RESULTS: The GAT-based pre- and post-operative IOP was 15.75 +/- 2.24 mm Hg, and 10.72 +/- 2.31 mm Hg, respectively, revealing the post-operative IOP to be significantly lower than the pre-operative IOP (p < 0.001). Among the 5 correction programs within Pentacam(R), Ehlers program showed little difference between pre- and post-operative IOP values (p = 0.228) and the post-operative correction value showed no significant difference with the pre-operative GAT value (p = 0.413). CONCLUSIONS: The Ehlers program is the most accurate among the 5 Pentacam(R) correction programs evaluated in the present study, and can be a useful tool for correcting the true IOP of patients which tends to be higher after corneal refractive surgery.
Key Words: Corneal refractive surgery;Correction program;Intraocular pressure;Pentacam(R)


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next