The Evaluation of the Difference between the Calculated Estimated Post-operative Refraction and the Real Post-operative Refraction for Five Types Intraocular Lenses.
Yong Eun Kim, Bum Jin Cho, Kang Suk Lee
1KangnamSL Eye Center, Korea. eyes421@eycos.co.kr
2Department of Ophthalmology, Ghil Medical Center, Gachon Medical School, Inchon, Korea.
3Department of Ophthalmology, College of Medicine, University of Ulsan, Asan Medical Center, Korea.
인공수정체 종류에 따른 예측굴절력과 실제굴절력의 차이
김용은 ( Yong Eun Kim ) , 조범진 ( Bum Jin Cho ) , 이강석 ( Kang Suk Lee )
Abstract
PURPOSE
To evaluate and compare the difference of predictive post-operative refraction preoperatively and real post-operative refraction among five types of intraocular lens. METHODS: We reviewed retrospectively 567 cataractous eyes that had undergone phacoemulsification or ECCE with posterior chamber intraocular lens (IOL) implantation by the same surgeon. Applied IOLs were AMO(R) Phacoflex(R) II SI40NB, Acrysof(R) MA60BM, Sensar(TM) AR40e, CeeOn(TM) 811B and CeeOn(TM) 720A. Prediction of post-operative refraction (predictive refraction) was calculated by the SRK/T formula with manufactured A constant. Post-operative manifest refraction (real refraction) was done at least 2 months postoperatively. We compared the difference between the predictive refraction and the real refraction by paired t-test. RESULTS: Total studied eyes were 390 eyes. In all groups, more myopic shift were observed than predicted. In AMO(R) Phacoflex(R) II SI40NB group, statistically significant difference was seen in postoperative manifest refraction over the predictive refraction as much as mean 0.46 diopter myopically (p<0.05, paired t-test), but others were not. In the AR40e implanted group, the error of predictive refraction was the smallest among groups. CONCLUSIONS: Prediction of postoperative refractive state was influenced by various clinical factors. Using a revised A constant, predictive error would be decreased. We recommend that every cataract surgeon had better have one's original A constant over each IOL and A constant may be revised when major surgical or biomedical measurement settings were changed.
Key Words: A constant;Predictive refraction;Refractive error;SI40NB


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next