Journal of the Korean Ophthalmological Society 2000;41(5):1164-1169.
Published online May 1, 2000.
Ultrasonic Determination of Axial Length in Pseudophakia.
Seong Bok Lee, Si Hwan Choi
Department of Ophthalmology, College of Medicine, Chungnam National University, Korea.
인공수정체안의 안축장 측정
이성복(Seong Bok Lee),최시환(Si Hwan Choi)
Abstract
To investigate the accuracy of ultrasonic measurement of axial length in pseudophakia, we measured axial length by ultrasonic scan at pre-and post-operatively.One hundred nine eyes of 97 patients were divided according to the type of intraocular lens.We compared pre-and post-operative axial lengths with Ocuscan[Alcon, U.S.A.]. The postoperative axial lengths were measured in phakic, aphakic, and pseudophakic mode.In pseudophakic mode, the velocity in IOL was set at the rate of 980, 2200 and 2718m/sec for silicone, acrylic, and PMMA lens respectively. The postoperative axial lengths in pseudophakic mode were closest to the preoperative values in all groups. In conclusion, we can get the closest values to preoperative axial lengths by setting the velocity in pseudophakic mode at the rate of 980, 2200 and 2718m/sec for silicone, acrylic and PMMA lens, respectively.
Key Words: Axial length, Calculation;Intraocular lens;Pseudophakia


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next