Journal of the Korean Ophthalmological Society 1999;40(6):1591-1598.
Published online June 1, 1999.
Transscleral Diode Laser Retinopexy in Retinal Detachment Surgery.
Boo Sup Oum, Jeong Young Kim
Department of Ophthalmology, Pusan National University Hospital.
망막박리수술에 있어서 Diode 레이저 경공막 망막유착술
임부섭(Boo Sup Oum),김정영(Jeong Young Kim)
Abstract
Authors performed transscleral diode laser retinopexy to evaluate its safety and efficacy in human eyes. Transscleral diode laser retinopexy was performed on 75 human eyes of rhegmatogenous retinal detachment. Power setting ranged from 400 to 1,000 mW, duration setting ranged from 0.5 to 1.0 second, and average 107 laser spots(range 15-443 laser spots) were applied. Seventy four of 75 eyes(98.7%) achieved anatomical success in reattachment after 6 months follow-up period. Complications of treatment encountered were minor Bruch's membrane rupture in 2 eyes, small retinal hemorrhage in one eye and mild scleral thermal burn in one eye. These complications did not affect adversely the surgical results. Among the 75 eyes, 21 eyes underwent diode laser choroidotomy for drainage of subretinal fluid with success. Transscleral diode laser retinopexy was considered as a safe and effective mean to obtain chorioretinal adhesion in retinal detachment surgery.
Key Words: Diode laser choroidotomy;Retinal detachment;Transscleral diode laser retinopexy


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
#1001, Jeokseon Hyundai BD
130, Sajik-ro, Jongno-gu, Seoul 03170, Korea
Tel: +82-2-2271-6603    Fax: +82-2-2277-5194    E-mail: kos@ijpnc.com                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next