The Effect of Machine Aging on the Measurements of Optical Coherency Tomography.
Min Sun Kim, Dae Yune Kim, Young Joon Jo, Jung Yeul Kim
1Department of Ophthalmology, Chungnam National University School of Medicine, Daejeon, Korea. kimjy@cnu.ac.kr
2Jung Eye Clinic, Geoje, Korea.
빛간섭단층촬영기의 노후화가 측정치에 미치는 영향
김민선1⋅김대윤2⋅조영준1⋅김정열1
충남대학교 의과대학 안과학교실1, 거제 정안과2
Abstract
PURPOSE
To evaluate the effect of instrument aging on optical coherence tomography (OCT) measurements. METHODS: Single eyes of 60 healthy participants (60 eyes) aged 20-30 years were examined twice using a 5-year-old OCT instrument and a new OCT instrument (Carl Zeiss Meditec, Dublin, CA, USA). The measurements and changes in signal strength were investigated using both instruments. RESULTS: The signal strengths of the new and aged instruments were 8.6 ± 0.8 and 5.0 ± 1.0, respectively, which was a statistically significant difference (p < 0.001). In addition, the central macular thicknesses (CMT) of the new and aged instruments were 201.1 ± 16.1 µm and 210.3 ± 16.0 µm, respectively. The thickness was significantly greater using the aged instrument (p < 0.001). Repeated measurements within the same eye were compared, and the difference in CMT was 3.2 using the new instrument and 10.5 using the aged instrument (p < 0.05). The intraclass correlation coefficient (ICC) was compared to evaluate the reproducibility of each instrument. The ICC values of nine areas of Early Treatment Diabetic Retinopathy Study in the aged instrument were 0.371-0.872, indicating low reproducibility. However, the new instrument showed high reproducibility with values of 0.806-0.947. CONCLUSIONS: Higher signal strength and lower CMT were observed using the new instrument compared to the aged instrument. Additionally, there were no differences in signal strength errors between the instruments. However, the error in CMT measured using the new instrument was significantly smaller compared to that using the aged instrument. Therefore, the effect of instrument aging should be accounted for in analyses of OCT measurements.
Key Words: Aging of optical coherence tomography (OCT);Central macular thickness;OCT measurements;Signal strength


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next