Comparison of Portable Tonometers and Goldmann Applanation Tonometer for Intraocular Pressure Measurement.
Hyun Ho Jung, Jung Won Park, Sang Woo Park
Department of Ophthalmology, Chonnam National University Medical School, Gwang, Korea. exo70@naver.com
휴대용 안압계와 골드만 압평안압계로 측정한 안압 비교
정현호⋅박정원⋅박상우
Department of Ophthalmology, Chonnam National University Medical School, Gwang, Korea
Abstract
PURPOSE
To compare the intraocular pressure (IOP) measured by portable rebound tonometer and TonoPen applanation tonometer with pressure measured by Goldmann applanation tonometer (GAT) and analyze the factors affecting IOP disagreement between tonometers. METHODS: In a prospective study of 463 eyes, IOP was measured with Icare Pro(R) rebound tonometer, TonoPen AVIA(R) applanation tonometer, and GAT. Bland-Altman plot, intraclass correlation coefficient, Pearson's correlation analysis, and multiple regression analysis were performed to evaluate the agreement of IOP measured by each tonometer and the factors affecting the measurements. RESULTS: The IOP values measured by Icare Pro(R) and TonoPen AVIA(R) were consistently higher than those measured by GAT, but showed no significant differences with those measured by GAT (p = 0.307 and 0.114, respectively). In Bland-Altman plot, the IOP values measured by Icare Pro(R) and TonoPen AVIA(R) exhibited excellent agreement with those measured by GAT. Both Icare Pro(R)/GAT and TonoPen AVIA(R)/GAT differences increased with younger age (p = 0.041 and 0.049, respectively) and higher central corneal thickness (p = 0.019 and 0.035, respectively). CONCLUSIONS: IOPs measured by portable Icare Pro(R) rebound tonometer and TonoPen AVIA(R) applanation tonometer were significantly correlated with IOP measured by GAT. Therefore, such instruments can be useful when measuring IOP with GAT is difficult. However, central corneal thickness and age should be considered when measuring IOP with portable tonometers.
Key Words: Goldmann applanation tonometer;Icare Pro(R) rebound tonometer;Intraocular pressure;TonoPen(R)


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next