Decentration, Tilt and Anterior Chamber Depth: Aspheric vs Spheric Acrylic Intraocular Lens.
Ja Young Lee, Seung Hee Lee, Sung Kun Chung
1Department of Ophthalmology, Seoul Adventist Hospital, Seoul, Korea.
2Department of Ophthalmology, St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, Korea. eyedoc@catholic.ac.kr
비구면 인공수정체와 구면 인공수정체의 중심이탈, 경사도 및 전방깊이의 비교
이자영1ㆍ이승희1ㆍ정성근2
Department of Ophthalmology, Seoul Adventist Hospital1, Seoul, Korea / Department of Ophthalmology, St. Mary’s Hospital, College of Medicine, The Catholic University of Korea2, Seoul, Korea
Abstract
PURPOSE
To compare the decentration, tilt and anterior chamber depth between aspheric AcrySof(R) IQ IOL (SN60WF) and spheric AcrySof(R) Natural IOL (SN60AT). METHODS: The subjects of this study consisted of 22 patients (26 eyes) using an aspheric AcrySof(R) IQ IOL and 26 patients (31 eyes) using spheric AcrySof(R) Natural IOL. All lenses were inserted into the capsular bag after 5 mm continuous curvilinear capsulorhexis and phacoemulsification. The decentration, tilt and anterior chamber depth of both IOLs were measured on the 1st, 30th and 60th postoperative day using an anterior eye segment analysis system (Scheimpflug camera, EAS-1000, Nidek, Japan). RESULTS: There was no statistically significant difference in the decentration between the IQ IOL and Natural IOL on the 1st (p=0.05), 30th (p=0.09) and 60th (p=0.06) postoperative day. There was a statistically significant difference in tilt between IQ IOL and Natural IOL on the 1st (p=0.000053) and 30th (p=0.018) postoperative day. However, there was no statistically significant difference in tilt on the 60th postoperative day (p=0.05). The anterior chamber depth of IQ IOL was decreased, but was not statistically significant on the 1st (p=0.10), 30th (p=0.07) or 60th (p=0.06) postoperative day. CONCLUSIONS: There were no significant differences between AcrySof(R) IQ IOL and AcrySof(R) Natural IOL in decentration, tilt or anterior chamber depth, showing that posterior aspheric surface causes little effect on the IOL stability.
Key Words: AcrySof(R) IQ IOL;AcrySof(R) natural IOL;Anterior chamber depth;Decentration;Tilt


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next