Journal of the Korean Ophthalmological Society 2005;46(4):676-681.
Published online April 30, 2005.
Clinical Application of Pressure Phosphene Tonometer: Comparison with Goldmann Applanation Tonometer in Normal Subjects.
Ho Yeop Yeom, Seung Jae Lee, Jong Hyun Lee, Chan Yun Kim, Young Jae Hong, Gong Je Seong
The Institute of Vision Research, Department of Ophthalmology, Yonsei University College of Medicine, Seoul, Korea. gjseong@yumc.yonsei.ac.kr
포스펜 안압계의 임상 적용 : 정상인에서 골드만 압평안압계와의 비교
염호엽,이승재,이종현,김찬윤,홍영재,성공제
The Institute of Vision Research, Department of Ophthalmology, Yonsei University College of Medicine, Seoul, Korea
Correspondence:  Ho-Yeop Yeom, M.D.
Abstract
PURPOSE
To assess the clinical usefulness of the pressure phosphene tonometer (Proview(R)) compared with the Goldmann applanation tonometer. METHODS: Ninety four eyes of 50 patients, aged from 3rd decade to 6th decade and with intraocular pressure (IOP) ranging from 9 to 21 mmHg, were measured by Proview(R) 5 times each. The average of these values was compared with the value measured with the Goldmann applanation tonometer. Twenty eyes of 10 patients, aged from 3rd decade to 4th decade, were trained repeatedly and measured again by the same method after 1 month. RESULTS: IOP measured by Proview(R) was statistically higher than that with the Goldmann applanation tonometer (p<0.05). The difference between the two measurements for patients in their 5th and 6th decades was more than in their 3rd and 4th decades. Nevertheless, the measurement difference decreased and the correlation coefficient increased between the devices in the repeatedly trained group (r=0.923). CONCLUSIONS: In measuring IOP, the pressure phosphene tonometer (Proview(R)) is not suitable in elderly or newly adapted patients, but it provides significant values in young and repeatedly trained group.
Key Words: Phosphene;Pressure phosphene tonometer;Proview(R)


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next