Comparison of Reproducibility between Exophthalmometer Readings Measured with Hertel and Naugle Exophthalmometer.
Jae Hyo Hwang, Sang Yeul Lee, Sung Joo Kim
1The Institute of Vision Research.
2Department of Ophthalmology, Yonsei University College of Medicine.
Hertel 안구돌출계와 Naugle 안구돌출계에 의한 안구돌출값의 재현성(reproducibility) 비교
황재효(Jae Hyo Hwang),이상열(Sang Yeul Lee),김성주(Sung Joo Kim)
Abstract
Hertel and Naugle exophthalmometers were used to compare the variability and reproducibility, produced by various observers. Five examiners measured fifty two eyes of twenty-six normal subjects each by using two different exophthalmometers, respectively. The average coefficient of variation among five measurements was 5.6% in the right eye, 5.2% in the left eye and 2.1% in base line by Hertel exophthalmometer. The average coefficient of variation measured by Naugle exophthalmometer was 1.9% in both eyes and 0.8% in pupillary distance. In conclusion, Naugle exophthalmometer showed less variability and better reproducibility; it is more appropriate for comparative exophthalmetry.
Key Words: Hertel exophthalmometer;Naugle exophthalmometer;Reproducibility