Journal of the Korean Ophthalmological Society 1990;31(6):739-745.
Published online June 1, 1990.
Gas-Liquid Chromamgraphic Analysis of Residual Ethylene Oxide and MMA Monomer in Various lntraocular Lenses.
In Sik Kim, Taek Jae Kim, Kyung Hwan Shyn
1Dept of Ophthalmology, Medical School, Chung Ang University , Korea.
2Chemical Analysis Laboratory, Korea Institute of Science and Technology, Korea.
각종 인공수정체의 잔류 EO gas 및 MMA monomer 측정에 관한 연구
신경환(Kyung Hwan Shyn),김인식(In Sik Kim),김택제(Taek Jae Kim)
Abstract
Recently, we performed an ezperi nlent to detect residual ethylene oxide(EO) gas and methylmethacrylate(MMA) monomer in the intraocular lenses from eleven different makers using gas-liquid chromatography. The residual EO gas over 25ppm was seen in 2 makers and also the MMA monomer over 2% was present in two makers. As a result, we recommend that the precise product quality-control of the IOLes is necessary especially for the lenses from foreign manufacturers.
Key Words: EO gas;IOL;GLC;MMA monomer;product quality-control