Comparison of Corneal Astigmatism and Higher-order Aberrations between Color Light-emitting Diode Topographer and Scheimpflug Imager.
Da Yeong Kim, Minji Ha, Rowoon Yi, Hyo Won Kim, So Hyang Chung
Department of Ophthalmology, College of Medicine, The Catholic University of Korea, Seoul St. Mary's Hospital, Seoul, Korea. chungsh@catholic.ac.kr
발광다이오드 각막형태분석기와 샤임플러그 전안부사진기의 각막난시 및 각막고위수차 비교
김다영 · 하민지 · 이로운 · 김효원 · 정소향
가톨릭대학교 의과대학 서울성모병원 안과
Correspondence:  So-Hyang Chung, MD, PhD
Email: chungsh@catholic.ac.kr
Received: 18 April 2019   • Revised: 28 May 2019   • Accepted: 7 June 2019
Abstract
PURPOSE
To compare corneal astigmatism, keratometry and corneal higher order aberrations between the light emitting diode corneal topography analyzer and Scheimpflug Imager. METHODS: This prospective study involved 45 patients (45 eyes) who visited Seoul St. Mary's hospital before cataract surgery from June 7, 2017, to August 2, 2017. For each eye, keratometry, astigmatism and its axis of cornea, higher-order aberrations were evaluated with a Scheimpflug Imager (Pentacam HR®, Oculus, Wetzlar, Germany) and a color-LED corneal topographer (Cassini®, i-Optics, Den Haag, The Netherlands). RESULTS: Astigmatism magnitude measured using Cassini® and Pentacam® showed no statistically differences but anterior and total astigmatic axes were significantly different, as measured by the two devices (p < 0.05). Anterior and total mean keratometry were statistically significantly different, as measured by the two devices (p < 0.05). J0 and J45 vectors of anterior and total cornea were statistically different (p < 0.05). In addition, Cassini® and Pentacam® showed discrepancies between total corneal astigmatism, total J0 and J45 vectors. Corneal anterior spherical aberration, vertical and horizontal coma, and oblique and horizontal trefoil aberrations were not statistically different between the two devices. CONCLUSIONS: Astigmatic axes obtained from the two devices based on different principles showed statistically significant differences. Astigmatism magnitude was not statistically different but showed a discrepancy between the two devices.
Key Words: Astigmatism;Corneal topography;Corneal wavefront aberration


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next