A Case of Fibrotic Obstruction of Scleral Orifice of Ex-PRESS Shunt.
Yung Hui Kim, Mi Sun Sung, Sang Woo Park
Department of Ophthalmology, Chonnam National University Medical School, Gwangju, Korea. exo70@naver.com
Ex-PRESS 방수유출장치 공막 유출부의 섬유성 폐쇄를 보인 1예
김영휘⋅성미선⋅박상우
전남대학교 의과대학 안과학교실
Correspondence:  Sang Woo Park,
Email: exo70@naver.com
Received: 23 July 2015   • Revised: 13 October 2015   • Accepted: 30 December 2015
Abstract
PURPOSE
To report a case of fibrotic obstruction of scleral orifice in Ex-PRESS shunt. CASE SUMMARY: A 57-year-old male presented with elevated intraocular pressure in his left eye. In the past, laser photocoagulation was performed at a local clinic due to retinal venous occlusion in his left eye. During observation, he was transferred to our hospital due to uncontrolled intraocular pressure despite antiglaucoma medications. He was diagnosed with neovascular glaucoma in the left eye. Since intravitreal injection of anti-vascular endothelial growth factor with maximal medical therapy did not lower the intraocular pressure, an Ex-PRESS shunt device was implanted. Two weeks postoperatively, the bleb was flat and diffuse with shallow anterior chamber and intraocular pressure was below 5 mm Hg. Therefore, we performed scleral flap revision and intraocular pressure was sustained between 10 and 15 mm Hg. Two months postoperatively, the patient experienced high intraocular pressure and no elevated bleb, thus we performed bleb revision. During the surgery, although we removed fibrotic adhesions between the conjunctiva and episclera using a 30-gauge needle, aqueous outflow was not observed. Therefore, we opened the scleral flap and found the scleral orifice of the Ex-PRESS shunt was obstructed by fibrous scar tissue. After scar tissue removal and achieving aqueous outflow through the scleral opening, intraocular pressure decreased to a satisfactory level. CONCLUSIONS: In patients with elevated intraocular pressure after implantation of the Ex-PRESS shunt, the scleral opening of the Ex-PRESS shunt should be examined for obstruction.
Key Words: Bleb revision;Ex-PRESS;Fibrotic obstruction;Scleral opening;Shunt


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next