Diagnostic Abilities to Detect Glaucomatous Abnormality Using Normal Retinal Thickness Measured by Optical Coherence Tomography.
Eun Jin Bae, Kyoung Lae Kim, Young Cheol Yoo
Department of Ophthalmology, Kangdong Sacred Heart Hospital, Hallym University College of Medicine, Seoul, Korea. demian7435@gmail.com
빛간섭단층촬영기로 측정한 국소지점별 망막두께 정상치를 활용한 녹내장성 시야감도저하 진단능력
배은진⋅김경래⋅유영철
Department of Ophthalmology, Kangdong Sacred Heart Hospital, Hallym University College of Medicine, Seoul, Korea
Abstract
PURPOSE
Recently, the introduction of spectral-domain optical coherence tomography (SD-OCT) has enabled measurement of retinal thickness in the posterior pole in 64 sectors. SD-OCT was used to evaluate the diagnostic effectiveness in detecting glaucomatous abnormality of visual field sensitivity. A normal value for retinal thickness was determined and then compared in corresponding local sectors. METHODS: Thirty healthy controls and 30 glaucoma subjects were evaluated. Macular thickness values from the 4 adjacent square cells in an 8 x 8 posterior pole retinal thickness map were averaged for a mean retinal thickness (MRT) value. A normative database was prepared using the data from the healthy eyes of this study to determine the diagnostic criteria for MRT. If the MRT value was <5% (Criteria A) or <1% (Criteria B) of the normative database, it was considered to be abnormal. The abnormalities of the MRT value for each diagnostic criteria were compared with the visual field sensitivity results in the corresponding positions. RESULTS: The concordance of abnormalities between MRT and visual field sensitivity at 16 measured points was low in both criteria A (Kappa value; -0.418~0.429) and B (Kappa value; -0.363~0.444). Based on the results of the visual field at each focal point, the sensitivities and specificities of MRT values using the 2 criteria ranged from 0% to 100%. CONCLUSIONS: In this study, MRT values showed low correlation and diagnostic ability to detect decreased sensitivity of the visual field in corresponding points, when customized criteria derived from a normative database were applied.
Key Words: Glaucoma;Retinal thickness;Spectral domain optical coherence tomography;Standard automated perimetry;Visual field sensitivity


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next