Formula Comparison for Intraocular Lens Power Calculation Using IOL Master and Ultrasound for the ZCB00 IOL.
Dong Hoon Shin, Dong Hui Lim, Ja Young You, Eui Sang Chung, Tae Young Chung
Department of Ophthalmology, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea. tychung@skku.edu
ZCB00 인공수정체의 부분결합간섭계와 초음파 방식의 술 후 굴절 예측치 차이
신동훈⋅임동희⋅유자영⋅정의상⋅정태영
Department of Ophthalmology, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea
Abstract
PURPOSE
To evaluation the accuracy of the IOL power calculation formulae measured by IOL Master(R) and applanation ultrasonography for the Tecnis ZCB00 IOL. METHODS: We performed a retrospective study of 170 eyes in 121 patients who underwent cataract surgery in our hospital with AMO Tecnis ZCB00 IOL.s. The SRK/T formula was used to predict the patient's implanted IOL power. Differences in the predicted refractive errors between IOL Master(R) and ultrasonography were analyzed and factors attributed to the differences were also analyzed. Three months after cataract surgery, mean numeric error and mean absolute error were analyzed. RESULTS: SRK/II and SRK/T formulas calculated using ultrasonography showed differences compared to the same formulas calculated using IOL Master(R), in which hyperopic shift was also demonstrated. No definite factor was attributed to the differences between the 2 methods. Although the 3 formulas of IOL Master(R) showed no significant difference in refractive errors, the SRK/T formula calculated using IOL Master(R) showed the least mean absolute and numeric errors. CONCLUSIONS: IOL Master(R) is considered more suitable when determining proper AMO Tecnis ZCB00 IOL power in cataract surgery. The hyperopic shift should be considered when calculating the IOL power using only ultrasonography.
Key Words: Intraocular lens power calculation;IOL Master(R);Ultrasound (A-scan);ZCB00


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next