Comparison of Rebound Tonometer, Non-Contact Tonometer, Goldmann Applanation Tonometer and the Relationship to Central Corneal Thickness.
Juhyang Lee, Mincheol Seong, Minho Kang, Heeyoon Cho, Yoonjung Lee
Department of Ophthalmology, Hanyang University Guri Hospital, Hanyang University College of Medicine, Guri, Korea. goddns76@hanmail.net
리바운드 안압계, 비접촉 안압계, 골드만 압평안압계의 비교 및 중심각막두께와의 상관성
이주향⋅성민철⋅강민호⋅조희윤⋅이윤정
Department of Ophthalmology, Hanyang University Guri Hospital, Hanyang University College of Medicine, Guri, Korea
Abstract
PURPOSE
To compare the level of accuracy of intraocular pressure (IOP) measurements of a rebound tonometer (RT) Icare(R), and non-contact tonometer (NCT), using Goldmann Applanation tonometer (GAT) as a reference to evaluate the influence of central corneal thickness (CCT) on IOP readings in the Korean population. METHODS: In a prospective study of 273 eyes, IOP was measured with RT, GAT, and NCT and compared to CCT measurements. Patients were assigned to one of 3 groups based on IOP measurements of GAT and 1 of 2 groups based on CCT. The comparison of the IOP values of RT, GAT, and NCT was performed between the IOP and CCT groups, and the differences among tonometers were evaluated. RESULTS: The RT showed statistically significant correlation with the GAT compared to the NCT. The CCT was related to RT measurements. The RT values compared to the GAT was underestimated in thin corneas and overestimated in thick corneas. CONCLUSIONS: There was a significant correlation between the RT and the GAT measurements compared with the NCT. However, RT is influenced by CCT and correlates less with GAT in low IOP ranges, suggesting that corneal thickness should be taken into consideration during such measurements.
Key Words: Central corneal thickness;Goldmann applanation tonometer;Non-contact tonometer;Rebound tonometer


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next