Intraocular Pressure Measurement with the Noncontact Tonometer and Rebound Tonometer through Plano Soft Contact Lenses.
Chang Mok Lee, Young Cheol Yu
Department of Ophthalmology, Kangdong Sacred Heart Hospital, Hallym University College of Medicine, Seoul, Korea. demian7435@gmail.com
편평 콘택트렌즈 착용안에서 비접촉 안압계와 리바운드 안압계를 이용한 안압 측정
이창목⋅유영철
Department of Ophthalmology, Kangdong Sacred Heart Hospital, Hallym University College of Medicine, Seoul, Korea
Abstract
PURPOSE
To investigate the accuracy of intraocular pressure (IOP) measurement through a plano soft contact lens (SCL) in situ by a noncontact tonometer (NCT) and a rebound tonometer (RBT). METHODS: The IOP of 66 eyes of 33 subjects with no ocular pathology was measured by NCT (tonometer CT-80, TOPCON, Japan) and RBT (iCare rebound tonometry(TM), Tiolat Oy, Finland). IOP measurement was repeated through plano SCL (ACUVUE OASYS(R), Johnson & Johnson Vision Care Inc., USA) in situ. Statistical analysis was performed using the paired t-test. RESULTS: Mean IOP measured by NCT was 13.10 +/- 2.52 mmHg without SCL and 12.95 +/- 2.56 mmHg with SCL. Mean IOP measured by RBT was 14.13 +/- 2.94 mmHg without SCL and 13.84 +/- 2.75 mmHg with SCL. No significant statistical differences were found between IOP measured with and without SCL (p=0.47, p=0.11 respectively). CONCLUSIONS: Reliable measurement of IOP by NCT and RBT can be achieved through plano SCL.
Key Words: Intraocular pressure;Noncontact tonometer;Rebound tonometer;Soft contact lens


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next