Journal of the Korean Ophthalmological Society 2012;53(1):1-10.
DOI: https://doi.org/10.3341/jkos.2012.53.1.1    Published online January 15, 2012.
Influences of Cylindrical Correction Amount of PRK on Accuracy of Geometric Corneal Center-Adjusted Ablation Centration.
Dae Young Park, Sang Bumm Lee
Department of Ophthalmology, Yeungnam University College of Medicine, Daegu, Korea. sbummlee@med.yu.ac.kr
굴절교정레이저각막절제술에서 각막기하중심-보정 연마중심의 정확도에 난시교정량이 미치는 영향
박대영⋅이상범
Department of Ophthalmology, Yeungnam University College of Medicine, Daegu, Korea
Abstract
PURPOSE
To evaluate the effect of the amount of cylindrical correction on the accuracy of geometric corneal center-adjusted ablation centration selected by the surgeon in advanced surface ablation-photorefractive keratectomy (ASA-PRK). METHODS: Ninety-five myopic eyes of 62 patients who underwent ASA-PRK were divided into three groups based on the amount of myopic cylindrical correction: Group 1 <-0.50 D, 22 eyes; Group 2 > or =-0.50 D and <-1.00 D, 45 eyes; Group 3 > or =-1.00 D, 28 eyes. The distances and distribution of the ablation centers from the entrance pupil center were analyzed by corneal topography at one month after ASA-PRK. Risk factors for decentration (>0.3 mm) were determined from patient-related factors such as gender, age, laterality, central corneal thickness, and order of procedure and surgery-related factors such as amount of spherical correction, ablation depth, vertical scale bar percentage, and learning effect of surgeon. RESULTS: Mean decentration was 0.26 +/- 0.15 mm (0.04 to 0.83 mm). There was no statistically significant difference in the amount of decentration among the three groups (p = 0.879). Superior-nasal displacement (53.7%) of the ablation center most frequently occurred after ASA-PRK. The amount of decentration among the three groups was not dependent on either patient-related or surgery-related factors. CONCLUSIONS: ASA-PRK using the surgeon-selected ablation center with adjustment toward the geometric corneal center was found to be highly accurate in ablation centration and did not influence the amount of ablation decentration regardless of the amount of myopic cylindrical correction.
Key Words: Ablation center;Cylindrical correction;Decentration;Geometric corneal center;Photorefractive keratectomy (PRK)
TOOLS
METRICS Graph View
  • 0 Crossref
  •  0 Scopus
  • 821 View
  • 11 Download
Related articles


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next