Journal of the Korean Ophthalmological Society 1995;36(9):1536-1540.
Published online September 1, 1995.
Examination of The Visual Field Change Results using The Humphrey Visual Field Analyzer and The Octopus Visual Field Analyzer in Early Primary Open Angle Glaucoma.
Woon Bong Jwa, Woo Chan Park, Sae Heun Rho
Department of Ophthalmology, College of Medicine, Dong-A University, Pusan, Korea.
조기 원발성 광우각형 녹내장에서 시행한 Humphrey 시야검사계 결과의 Octopus 시야검사계에 의한 검정
좌운봉(Woon Bong Jwa),박우찬(Woo Chan Park),노세현(Sae Heun Rho)
Abstract
The Humphrey Visual Field Analyzer and The Octopus Visual Field Analyzer in Early Primary Open Angle Glaucoma Visual field measurements are useful tool not only for diagnosis but also for management ealuation of early glaucoma. To know the relationship between the results of visual field which were performed by the Humphrey and the Octopus, the 25 disc data which have shown the general reduction of sensitivity on two continuous separate sessions in Humphrey viual field analyzer were transmitted into the Peridata Standard Printout Mode of GIX program on the Octopus 1-2-3. We compared the aspect of Bebie curve of Octopus 1-2-3 and the difference of the global indices between Humphrey and Octopus in the basically input data of age related standard sensitivity of visual field. The general reduction of sensitivity in Humphrey visual field analyzer showed the sensitivity loss in Bebie curve of Octopus 1-2-3(p<5%). In the statistical analysis of Global indices between the two field analyzers there were no differences in the MD and PSD(p>0.05) but were difference the CPSD and SF(p<0.05). In the comparison of the correlation on global indices between the two field analysis. MD and PSD were closely correlated r=0.657, PSD r=0.697, p<0.001), CPSD and SF were statistically correlated(p<0.001) but had lower correlation coefficients(CPSD r=0.478, SF r=0.389). Thus we conclude that there are little differences between the results of two field analyzers.
Key Words: Bebie curve;General reduction of sensitivity;Global index;Humphrey visual field analyzer;Octopus visual field analyzer
TOOLS
METRICS Graph View
  • 417 View
  • 6 Download
Related articles


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next