Comparison of Colvardpupillometer, ORBScan II and Sirius in Determining Pupil Size for Refractive Surgery.
Ji Sun Baek, Ji Hyun Park, Eun Seok Yoo, Young A Kwon, Sang Wroul Song, Byoung Yeop Kim, Jae Lim Chung
1Department of Ophthalmology, Konyang University, Kim's Eye Hospital, Seoul, Korea. jlchung@kimeye.com
2Daesung Yonsei Eye Clinic, Bucheon, Korea.
3Minduelre Eye Clinic, Daejeon, Korea.
굴절교정수술시 Colvard 동공측정기, ORBScanⅡ, Sirius로 측정한 동공크기의 비교
백지선1⋅박지현2⋅유은석3⋅권영아1⋅송상률1⋅김병엽1⋅정재림1
Department of Ophthalmology, Konyang University, Kim’s Eye Hospital1, Seoul, Korea Daesung Yonsei Eye Clinic2, Bucheon, Korea Minduelre Eye Clinic3, Daejeon, Korea
Abstract
PURPOSE
To compare pupil size measured by Colvard(R)pupillometer, ORBScan II(R), Sirius(R) and analyze correlation between pupil size and night vision disturbance after refractive surgery according to measuring methods. METHODS: Two hundred eyes of one hundred patients who underwent refractive surgery were retrospectively evaluated. Pupil size was measured with Colvard(R)pupillometer, Sirius(R) in the scotopic light condition and with ORBScan II(R), Sirius(R) in the photopic light condition. Patients filled out questionnaire about night vision disturbance at postoperative three months. Correlation between pupil size according to measuring methods and night vision disturbance was evaluated. RESULTS: The mean age of patient was 26 years and number of male patients was 36. In scotopic light condition, pupil size measured with Colvard(R)pupillometer (6.76 mm) was significantly larger than that of Sirius(R) (6.53 mm) (p < 0.001). In photopic light condition, pupil size measured with ORBScan II(R) (3.98 mm) was significantly smaller than that of Sirius(R) (4.41 mm) (p < 0.001). Night vision disturbance were correlated with pupil size and Sirius(R) in the scotopic light condition had strongest correlation among three measuring method (r = 0.44). CONCLUSIONS: Sirius(R) tends to measure pupil size smaller than Colvard(R)pupillometer in the scotopic light condition. The correlation between night vision disturbance and pupil size was highest in Sirius(R).
Key Words: Colvard(R)pupillometer;Night vision disturbance;ORBScan II(R);Pupil size;Sirius(R)


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next